Class II — GE Healthcare, LLC
| Recalling firm | GE HealthCare Technologies Inc. |
|---|---|
| Registered facility | View this company's 50 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person. |
| Product | GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000. |
| Quantity | 667 devices |
| Distribution | Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution. Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Malaysia, New Z… |
| Initiated | 2022-05-27 |
| Classified | 2022-07-20 |
| Recall number | Z-1427-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — GE Healthcare, LLC (Z-1427-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90422/
Alerts for GE HealthCare Technologies Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.