RegWardenFDA facility compliance records
Recall

Class II — Menarini Silicon Biosystems

Recalling firmMenarini Silicon Biosystems Inc.
Registered facilityMenarini Silicon Biosystems Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonHigh number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples
ProductCELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001
Quantity484 units
DistributionCA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,
Initiated2022-05-20
Classified2022-07-07
Recall numberZ-1345-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Menarini Silicon Biosystems (Z-1345-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90417/

Not open yet

Alerts for Menarini Silicon Biosystems Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.