Class II — Covidien, LP
| Recalling firm | Covidien llc |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The affected staplers have the potential for the staple guide to not be securely attached to the instrument. The issue could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. |
| Product | EEA Autosuture Circular Stapler with DST Series Technology, 25mm. Model Numbers EEA25, EEAXL25, EEA2535, and EEAXL2535. |
| Quantity | 433,560 |
| Distribution | Worldwide distribution - US Nationwide. |
| Initiated | 2022-05-13 |
| Classified | 2022-07-01 |
| Recall number | Z-1322-2022 |
| Initiation | Voluntary: Firm initiated |