Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm |
| Product | Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700 |
| Quantity | 13 units |
| Distribution | US Nationwide distribution. |
| Initiated | 2022-04-11 |
| Classified | 2022-06-07 |
| Recall number | Z-1246-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1246-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90184/