Class II — Shanghai United Imaging Healthcare Co., Ltd.

Recalling firmShanghai United Imaging Healthcare Co.,Ltd.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonX-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure.
ProductDigital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h
Quantity4
DistributionUS Nationwide distribution in the state of TX.
Initiated2019-07-30
Classified2022-05-12
Terminated2022-05-17
Recall numberZ-1075-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Shanghai United Imaging Healthcare Co., Ltd. (Z-1075-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90104/

Not open yet

Alerts for Shanghai United Imaging Healthcare Co.,Ltd.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.