RegWardenFDA facility compliance records
Recall

Class II — GE Healthcare, LLC

Recalling firmGE HealthCare Technologies Inc.
Registered facilityView this company's 50 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonIf the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.
ProductGE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.
Quantity10,445 units
DistributionWorldwide distribution - US nationwide, including Puerto Rico. There was also government/military distribution. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cote d'lvoire, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, Germany, Ghana, Hong Kong, Hunga…
Initiated2022-01-28
Classified2022-03-16
Recall numberZ-0808-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GE Healthcare, LLC (Z-0808-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89576/

Not open yet

Alerts for GE HealthCare Technologies Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.