Class II — Siemens Medical Solutions USA, Inc.
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study documentation during a high risk procedure, such as a stress echo exam. |
| Product | Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system intended for the following applications: Vascular, Abdominal, Neonatal, Echocardiographical, Pediatric, Small Parts, Musculoskeletal, Obstetrical, Gyneological, Intraoperative, and Urological applications using different ultrasound transducers for different applications. |
| Quantity | 2846 |
| Distribution | Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra, Argentina, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kosovo, Ku… |
| Initiated | 2022-01-19 |
| Classified | 2022-03-11 |
| Recall number | Z-0748-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc. (Z-0748-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89499/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.