Class II — Qiagen Sciences LLC
| Recalling firm | QIAGEN Sciences LLC |
|---|---|
| Registered facility | QIAGEN (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur. |
| Product | QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223 |
| Quantity | 376 kits |
| Distribution | Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution. |
| Initiated | 2021-11-08 |
| Classified | 2022-01-05 |
| Terminated | 2023-03-29 |
| Recall number | Z-0455-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Qiagen Sciences LLC (Z-0455-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89130/
Alerts for QIAGEN Sciences LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.