RegWardenFDA facility compliance records
Recall

Class II — The Binding Site Group, Ltd.

Recalling firmTHE BINDING SITE, INC.
Registered facilityTHE BINDING SITE, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDue to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
ProductFreelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
Quantity2,254 kits
DistributionU.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada
Initiated2021-10-15
Classified2021-11-19
Terminated2025-02-04
Recall numberZ-0274-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — The Binding Site Group, Ltd. (Z-0274-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88921/

Not open yet

Alerts for THE BINDING SITE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.