RegWardenFDA facility compliance records
Recall

Class II — FujiFilm Healthcare Americas Corporation

Recalling firmFUJIFILM Healthcare Americas Corporation
Registered facilityFUJIFILM Healthcare Americas Corporation (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDiagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
ProductLisendo 880 Ultrasound Software Version: V1.0 through V4.1.3
QuantityN/A
DistributionNationwide
Initiated2021-10-14
Classified2021-11-02
Recall numberZ-0191-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — FujiFilm Healthcare Americas Corporation (Z-0191-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88840/

Not open yet

Alerts for FUJIFILM Healthcare Americas Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.