RegWardenFDA facility compliance records
Recall

Class I — MEDTECH SAS

Recalling firmMEDTECH S.A.S
Registered facilityMEDTECH SAS (the firm's only registered facility)
ClassificationClass I
StatusOngoing
ReasonThe firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect trajectory could result in serious injury or death if undetected during surgery.
ProductROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide standard surgical instruments during brain and spine surgeries.
Quantity119 (US); 55 (OUS)
DistributionUS Nationwide distribution and International Distribution to countries of: Australia, Taiwan, China, Thailand, Japan, India, South Korea, and the Netherlands.
Initiated2021-09-22
Classified2021-10-25
Recall numberZ-0118-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — MEDTECH SAS (Z-0118-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88746/

Not open yet

Alerts for MEDTECH S.A.S

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.