Class II — Datascope Corp.
| Recalling firm | Datascope Corp. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions. |
| Product | Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85 |
| Quantity | 3969 total (US); 3550 total (OUS) |
| Distribution | Nationwide domestic distribution. Foreign distribution worldwide. |
| Initiated | 2021-09-08 |
| Classified | 2021-09-30 |
| Recall number | Z-2616-2021 |
| Initiation | Voluntary: Firm initiated |