Class II — Beckman Coulter Inc.
| Recalling firm | BECKMAN COULTER, INC. |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples. |
| Product | Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK |
| Quantity | 293,640 kits |
| Distribution | U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, China, Croatia, Curaao, Czech Republic, Denmark, Egypt, … |
| Initiated | 2021-08-09 |
| Classified | 2021-10-05 |
| Recall number | Z-0032-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Beckman Coulter Inc. (Z-0032-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88559/
Alerts for BECKMAN COULTER, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?