RegWardenFDA facility compliance records
Recall

Class II — Aspen Surgical Products, Inc.

Recalling firmAspen Surgical Products
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonAffected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.
ProductJessie Sterile Arm Suspension Kit, Product Code 711300, (previous product code 100401), Sample Kit S711300
Quantity1429
DistributionUS nationwide distribution. No foreign distribution.
Initiated2021-07-23
Classified2021-08-30
Terminated2022-07-26
Recall numberZ-2388-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Aspen Surgical Products, Inc. (Z-2388-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88476/

Not open yet

Alerts for Aspen Surgical Products

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.