Class II — Roche Diagnostics Operations, Inc.
| Recalling firm | Roche Diagnostics Operations Inc. |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument. |
| Product | cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001 |
| Quantity | 2058 licenses worldwide, 79 licenses in US |
| Distribution | US Nationwide distribution |
| Initiated | 2021-06-24 |
| Classified | 2021-09-10 |
| Terminated | 2023-07-21 |
| Recall number | Z-2444-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-2444-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88398/
Alerts for Roche Diagnostics Operations Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?