RegWardenFDA facility compliance records
Recall

Class II — Oakworks Inc

Recalling firmOakworks, INC.
Registered facilityOAKWORKS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWeld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional
ProductVascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219
Quantity55 units
DistributionNationwide
Initiated2021-06-29
Classified2021-08-25
Terminated2023-10-05
Recall numberZ-2349-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Oakworks Inc (Z-2349-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88365/