RegWardenFDA facility compliance records
Recall

Class II — Terumo Cardiovascular Systems Corporation

Recalling firmTerumo Cardiovascular Systems Corp.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
Reasonintermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
ProductTerumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
Quantity1 unit OUS
DistributionInternational distribution in the country of Panama.
Initiated2021-06-08
Classified2021-08-26
Terminated2022-06-28
Recall numberZ-2350-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Terumo Cardiovascular Systems Corporation (Z-2350-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88350/

Not open yet

Alerts for Terumo Cardiovascular Systems Corp.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.