Class II — Covidien Llc
| Recalling firm | Covidien llc |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Potential for a manufacturing assembly error-may result in a non-functional audio indicator of device activation or a non-functional device, may cause unintended thermal burn, hemorrhage, or tissue damage and a delay of treatment while an alternate device is located |
| Product | Covidien Sonicision Reusable Generator-converts electrical power from the battery pack to ultrasonic energy to drive the active blade. Product Number: SCGAA |
| Quantity | 96 units |
| Distribution | US Nationwide Distribution: AL, AZ, CT, GA, IN, MA, MD, MN, NY, OH, TX, VA and WA. |
| Initiated | 2021-06-23 |
| Classified | 2021-08-10 |
| Terminated | 2025-01-28 |
| Recall number | Z-2240-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Covidien Llc (Z-2240-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88271/