RegWardenFDA facility compliance records
Recall

Class II — GE Healthcare, LLC

Recalling firmGE HealthCare Technologies Inc.
Registered facilityView this company's 50 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusCompleted
ReasonGE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.
ProductRevolution CT
Quantity197 (US = 40; OUS = 157)
DistributionWorldwide distribution - US Nationwide
Initiated2021-03-19
Classified2021-07-08
Recall numberZ-2004-2021
InitiationFDA Mandated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GE Healthcare, LLC (Z-2004-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88150/