Class II — Omnia SRL

Recalling firmOmnia SRL
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonOmnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.
ProductOmnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32.Z2034.00, 32.Z2036.00, 32.Z2038.00, 32.Z2040.00, 72.T1950.00, 72.T1951.00, 72.T1952.00, 72.T1953.0…
Quantity276,901 units
DistributionDomestic: AK, IL, PA, OK, TX, UT, NY, CA. Foreign: Australia, Austria, Belgium, Bulgaria, Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Lebanon, Netherlands, New Zealand, Poland, Russia, Slovenia, Spain, Sweden, Switzerland, United Kingdom, Uzbekistan, Canada, Malta,
Initiated2021-03-30
Classified2021-04-29
Terminated2023-11-08
Recall numberZ-1518-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Omnia SRL (Z-1518-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87680/

Not open yet

Alerts for Omnia SRL

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.