Class II — Greiner Bio-One North America, Inc.
| Recalling firm | GREINER BIO-ONE NA |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | VACUETTE® Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays. |
| Product | VACUETTE® Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing. |
| Quantity | 1,049,700 units |
| Distribution | US Nationwide distribution in the states of KY, TX, LA, OH, AR, IL, SC, PA, FL, NC, AZ, CA, MI, NE, KS, MI, MO, NY, and NJ. |
| Initiated | 2021-03-29 |
| Classified | 2021-05-06 |
| Terminated | 2023-10-10 |
| Recall number | Z-1534-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Greiner Bio-One North America, Inc. (Z-1534-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87660/
Alerts for GREINER BIO-ONE NA
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?