RegWardenFDA facility compliance records
Recall

Class II — Covidien Llc

Recalling firmCovidien llc
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPotential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch
ProductCovidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia stapling system. Staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. Product Number: SIGTRSB60AMT
Quantity222 units
DistributionAL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX
Initiated2021-03-23
Classified2021-04-13
Terminated2024-05-15
Recall numberZ-1419-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Covidien Llc (Z-1419-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87618/

Not open yet

Alerts for Covidien llc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.