Class II — Hologic, Inc
| Recalling firm | HOLOGIC, INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient results. |
| Product | Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary consumable to the Panther Fusion System that is used to contain the PCR and RT-PCR steps of the assay processing. |
| Quantity | 1355 kits |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CT, DC, FL, GA, HI, IL, IN, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NY, OH, OR, PA, TX, UT, VA, VT, WA, WI, WV and the countries of Austria (AT), Belgium (BE), France (FR), Germany (DE), Great Britain (GB), Italy (IT) Netherlands (NL), Spain (ES), Switzerland (CH),, Canada (CA), Hong Kong (HK), Australia (AU), New… |
| Initiated | 2021-02-26 |
| Classified | 2021-04-07 |
| Recall number | Z-1369-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Hologic, Inc (Z-1369-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87557/
Alerts for HOLOGIC, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.