RegWardenFDA facility compliance records
Recall

Class II — Insulet Corporation

Recalling firmINSULET CORPORATION
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonCertain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended
ProductOmnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 000319 Distributed to: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland
Quantity21,723 guides
DistributionCanada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland
Initiated2021-02-01
Classified2021-03-10
Terminated2023-02-08
Recall numberZ-1203-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Insulet Corporation (Z-1203-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87305/