RegWardenFDA facility compliance records
Recall

Class II — Steris Corporation

Recalling firmSTERIS Corporation
Registered facilityKey Surgical, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.
ProductReliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
Quantity3405
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Bulgaria, Canada, Chile, China, Ecuador, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of Malaysia, Mexico, New Zealand, Panama, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela.
Initiated2020-12-11
Classified2021-01-15
Terminated2022-06-21
Recall numberZ-0876-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Steris Corporation (Z-0876-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86937/

Not open yet

Alerts for STERIS Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.