RegWardenFDA facility compliance records
Recall

Class II — Boston Scientific Corporation

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThis is a follow-up to the EKOS Field Safety Notice dated April 30, 2019. The Radio Frequency (RF) board has been redesigned to prevent occasional internal board failures at lower temperatures. All EKOS CU 4.0 consoles containing the old RF board design will be replaced with the newly designed CU 4.0 consoles.
ProductEKOS Control System (CU) 4.0. UPN 600-40500 The EkoSonic MACH4 Endovascular (EKOS) System consists of a single use Infusion Catheter and Ultrasonic Core, and a reusable Control Unit 4.0. The device delivers the physician-specified fluids and ultrasound to the intravascular treatment site. The reusable Control Unit provides power to the device and provides the user interface for operator control. A reusable, non-sterile connector interface cable (CIC) connects the Control Unit to the Ultrasonic Core and Infusion Catheter.
Quantity90 (OUS)
DistributionNo US distribution. Foreign distribution to Latin America and EU.
Initiated2020-11-03
Classified2020-12-17
Recall numberZ-0637-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Boston Scientific Corporation (Z-0637-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86847/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.