Class II — Covidien Llc
| Recalling firm | Covidien llc |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm. |
| Product | Capnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused. |
| Quantity | 3217 |
| Distribution | Worldwide distribution - US Nationwide distribution. |
| Initiated | 2020-11-16 |
| Classified | 2020-12-01 |
| Recall number | Z-0515-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Covidien Llc (Z-0515-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86829/
Alerts for Covidien llc
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?