RegWardenFDA facility compliance records
Recall

Class II — Invacare Corporation

Recalling firmInvacare Corporation dba Invacare America
Registered facilityView this company's 7 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBreakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.
ProductInvacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5
Quantity384,767 units
DistributionWorldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands
Initiated2020-10-30
Classified2020-12-01
Terminated2024-12-31
Recall numberZ-0514-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Invacare Corporation (Z-0514-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86604/