Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or arthroscope in the correct orientation.
Product5.5MM Arthrogarde Hip Access Cannula
Quantity143 units
DistributionDomestic Distribution: CA, CO, CT, GA, IA, IL, IN, MD, MO, NC, NY, OH, PA, SC, TN, TX, WA, WI, WV. International Distribution: AU, BR, CA, CL, CN, CO, DE, ES, GB, HK, KR, PA, PT, TW.
Initiated2020-09-21
Classified2020-10-27
Terminated2024-03-22
Recall numberZ-0301-2021
InitiationVoluntary: Firm initiated