Class I — Flexicare Medical Ltd.

Recalling firmFLEXICARE MEDICAL LTD.
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonWhen the video stylet cuff is over-inflated there is potential for occlusion of the corrugated section of the ETT, and increased resistance to gas flow and difficulty in ventilating the patient.
ProductProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U/ 7.0, 038-990-065U/ 6.5 *Note: product list updated per 12/23/2020 letter to customers
Quantity2087
DistributionUS nationwide distribution.
Initiated2020-09-22
Classified2020-11-26
Terminated2024-07-29
Recall numberZ-0497-2021
InitiationVoluntary: Firm initiated