RegWardenFDA facility compliance records
Recall

Class II — Siemens Healthcare Diagnostics, Inc

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSiemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.
ProductAtellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)3247…
QuantityAtellica IM H2n Assay =135 kits; Atellica IM H2n Calibrator = 46 kits; ADVIA Centaur Assay = 1,823 kits; ADVIA Centaur calibrator = 161 kits
DistributionUS Nationwide distribution.
Initiated2020-08-25
Classified2020-10-16
Terminated2023-03-10
Recall numberZ-0115-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc (Z-0115-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86383/