Recall

Class II — Zeus Scientific, Inc.

Recalling firmZEUS SCIENTIFIC LLC
Registered facilityZEUS SCIENTIFIC LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonFDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
ProductZEUS ELISA Parvovirus B19 IgM Test System, UDI 00845533001755 - Product Usage: intended for the qualitative detection of IgM class antibodies to human Parvovirus B19 in human serum including women of childbearing age where there is a suspicion of exposure to human Parvovirus B19.
Quantity74
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, CT, IN, NJ, TX and the country of Canada.
Initiated2019-08-12
Classified2020-09-29
Terminated2022-05-20
Recall numberZ-3026-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zeus Scientific, Inc. (Z-3026-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86364/

Not open yet

Alerts for ZEUS SCIENTIFIC LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.