Class II — Roche Diagnostics Operations, Inc.
| Recalling firm | Roche Diagnostics Operations Inc. |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated. |
| Product | Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190 |
| Quantity | 11,870 reagent kits |
| Distribution | US Nationwide |
| Initiated | 2020-08-20 |
| Classified | 2020-09-17 |
| Terminated | 2021-10-22 |
| Recall number | Z-2981-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)