RegWardenFDA facility compliance records
Recall

Class II — Cook Inc.

Recalling firmCOOK, INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonMay contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU
ProductZenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms Reference Part Number (RPN): ZIMB-22-108 ZIMB-22-118 ZIMB-22-128 ZIMB-22-70 ZIMB-22-84 ZIMB-22-98 ZIMB-24-108 ZIMB-24-118 ZIMB-24-128 ZIMB-24-70 ZIMB-24-84 ZIMB-24-98 ZIMB-26-108 ZIMB-26-118 ZIMB-26-128 ZIMB-26-70 ZIMB-26-84 ZIMB-26-98 ZIMB-28-108 ZIMB-28-118 ZIMB-28-128 ZIMB-28-70 ZIMB-28-84 ZIMB-28-98 ZIMB-30-108 ZIMB-30-118 ZIMB-30-128 ZIMB-30-70 ZIMB-30-84 ZIMB-30-98 ZIMB-32-108 ZIMB-32-108 …
Quantity684 units
DistributionForeign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPAA FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong JAPAN NEW CALEDONIA New Zealand Thailand
Initiated2020-08-11
Classified2020-09-11
Terminated2021-07-28
Recall numberZ-2964-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cook Inc. (Z-2964-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86242/