Class II — Biosense Webster (israel), Ltd.

Recalling firmBIOSENSE WEBSTER, INC.
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSoftware defect may result in disapperance of tag sites during recalculation and lead to additional ablation sites. This may lead to prolongation of the procedure and, in extremely rare circumstances, cardiac perforation.
ProductCARTO 3 System (Software Version V7.1.80), CARTO VISITAG Module - Product Usage: FGThe intended use of the system is catheter-based cardiac electrophysiological (EP) procedures. The system provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure, Product Number: -5400-00, UDI: 10846835000870
QuantityN/A
DistributionWorldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI *Three (3) U.S. government consignees. The countries of Australia, Austria, Belgium, Chile, China, Colombia, Czechia, Denmark, Finland, France, Ge…
Initiated2020-06-22
Classified2020-08-24
Terminated2022-01-06
Recall numberZ-2880-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Biosense Webster (israel), Ltd. (Z-2880-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85910/

Not open yet

Alerts for BIOSENSE WEBSTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.