Class II — Shanghai United Imaging Healthcare Co., Ltd.

Recalling firmShanghai United Imaging Healthcare Co.,Ltd.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonTwo issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.
ProductuCT 760 System, Computed Tomography X-ray System
Quantity5 units
DistributionDistributed to two customers. Both customers located in Houston, Texas.
Initiated2020-05-08
Classified2020-07-06
Terminated2022-01-20
Recall numberZ-2522-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Shanghai United Imaging Healthcare Co., Ltd. (Z-2522-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85804/

Not open yet

Alerts for Shanghai United Imaging Healthcare Co.,Ltd.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.