Class II — GE Healthcare, LLC
| Recalling firm | GE HealthCare Technologies Inc. |
|---|---|
| Registered facility | View this company's 50 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor. |
| Product | CARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro, Model Numbers: 2042084-001 2094504-001 2042084-001-046050 2042084-001-208621 2042084-001-045915 2042084-001-215164 2042084-001-078474 2042084-001-053034 2042084-001-215165 2042084-001-079462 2042084-001-047389 2042084-001-112198 2042084-001-041673 2042084-001-227238 2042084-001-071564 2042084-001-503824 2042084-001-119839 2042084-001-569269 2042084-001-041674 2042084-001-047549 2042084-001-514651 2042084-001-163572 2042084-001-583721 2042084-0… |
| Quantity | 89,109 units |
| Distribution | Worldwide Distribution |
| Initiated | 2020-05-18 |
| Classified | 2020-06-23 |
| Recall number | Z-2424-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — GE Healthcare, LLC (Z-2424-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85765/
Alerts for GE HealthCare Technologies Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.