RegWardenFDA facility compliance records
Recall

Class II — Carefusion 2200 Inc

Recalling firmCAREFUSION
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBecton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.
ProductMueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.
Quantity358 units
DistributionUS Nationwide distribution including in the states of CA, FL, VA and KY.
Initiated2020-05-18
Classified2020-06-18
Terminated2021-04-25
Recall numberZ-2400-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Carefusion 2200 Inc (Z-2400-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85747/