Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. |
| Product | Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194 |
| Quantity | 20 |
| Distribution | US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico. |
| Initiated | 2020-04-17 |
| Classified | 2020-05-26 |
| Terminated | 2020-08-25 |
| Recall number | Z-2128-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-2128-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85572/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?