Recall

Class II — Qiagen Sciences LLC

Recalling firmQIAGEN Sciences LLC
Registered facilityQIAGEN (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.
Producttherascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121
Quantity530 kits within expiry
DistributionUS Nationwide.
Initiated2020-04-08
Classified2020-06-05
Terminated2021-06-03
Recall numberZ-2290-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Qiagen Sciences LLC (Z-2290-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85565/