RegWardenFDA facility compliance records
Recall

Class II — The Binding Site Group, Ltd.

Recalling firmTHE BINDING SITE, INC.
Registered facilityTHE BINDING SITE, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonAssigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.
ProductOptilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)
Quantity478 kits
DistributionUS: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None
Initiated2019-10-07
Classified2020-05-22
Terminated2022-01-11
Recall numberZ-2110-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — The Binding Site Group, Ltd. (Z-2110-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85366/

Not open yet

Alerts for THE BINDING SITE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.