Recall

Class II — Haemonetics Corporation

Recalling firmHaemonetics Corporation
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSoftware defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.
ProductTEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies
Quantity88 units
DistributionUS Nationwide distribution.
Initiated2020-02-19
Classified2020-03-31
Terminated2021-06-21
Recall numberZ-1617-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Haemonetics Corporation (Z-1617-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85120/

Not open yet

Alerts for Haemonetics Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.