Class II — Intuitive Surgical, Inc.
| Recalling firm | Intuitive Surgical, Inc. |
|---|---|
| Registered facility | View this company's 41 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images. |
| Product | da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures. |
| Quantity | 1 system |
| Distribution | US Nationwide distribution including in the state of Washington. |
| Initiated | 2019-06-17 |
| Classified | 2020-10-23 |
| Terminated | 2022-07-18 |
| Recall number | Z-0283-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-0283-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84963/
Alerts for Intuitive Surgical, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.