Class II — The Binding Site Group, Ltd.

Recalling firmTHE BINDING SITE, INC.
Registered facilityTHE BINDING SITE, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonIt has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal results which may lead to unnecessary additional serological testing.
ProductRheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A. , UDI: 05051700018319
Quantity30 kits
DistributionDistributed in CA.
Initiated2019-11-06
Classified2020-03-13
Terminated2022-09-08
Recall numberZ-1500-2020
InitiationVoluntary: Firm initiated