RegWardenFDA facility compliance records
Recall

Class III — Theralase Inc.

Recalling firmTHERALASE INC.
Registered facilityTheralase Technologies Inc. (the firm's only registered facility)
ClassificationClass III
StatusTerminated
ReasonPower Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
ProductTLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002
Quantity202 units
DistributionU.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Initiated2019-09-26
Classified2020-02-24
Terminated2020-12-03
Recall numberZ-1348-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class III recall — Theralase Inc. (Z-1348-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84842/