Class II — Tosoh Bioscience Inc

Recalling firmTOSOH BIOSCIENCE
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.
ProductAIA-360 Automated Immunoassay Analyzer, Product Code 019945
Quantity12
DistributionDistributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay
Initiated2020-01-10
Classified2020-03-02
Terminated2020-09-10
Recall numberZ-1420-2020
InitiationVoluntary: Firm initiated