Class I — Medtronic Inc.
| Recalling firm | Medtronic |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia. |
| Product | Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MMT-1782 |
| Distribution | US:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Republic, Finland, Switzerland, Slovakia, Ireland, Poland, S. Africa, Greece, Hungary, Saudi Arabia, Unit… |
| Initiated | 2019-11-21 |
| Classified | 2020-02-07 |
| Recall number | Z-0956-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Medtronic Inc. (Z-0956-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84430/
Alerts for Medtronic
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.