Class II — Nucletron BV

Recalling firmNucletron B.V.
Registered facilityNUCLETRON B.V. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA component failure may lead to treatment interruption or incorrect source positioning.
ProductThere is a complaint regarding internal check cable guide mechanism failure for the Nucletron Flexitron brachytherapy afterloader system. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged. This issue could cause treatment interruption or in a very unlikely scenario, incorrect source positioning and radiation treatment errors.
Quantity383
DistributionDistribution in United States and worldwide
Initiated2019-11-18
Classified2020-01-02
Terminated2023-03-28
Recall numberZ-0772-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Nucletron BV (Z-0772-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84371/

Not open yet

Alerts for Nucletron B.V.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.