RegWardenFDA facility compliance records
Recall

Class II — Trilliant Surgical, LLC

Recalling firmTrilliant Surgical
Registered facilityTrilliant Surgical LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.
ProductGridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.
Quantity20 units
DistributionUS Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None
Initiated2019-03-14
Classified2020-01-10
Terminated2022-03-16
Recall numberZ-0799-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Trilliant Surgical, LLC (Z-0799-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84350/