Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonOn the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.
ProductAtellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens.
Quantity1528
DistributionThe products were distributed to the following foreign countries: Argentina, Belgium, Brazil, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland, and Vietnam. No affected products distributed in the United States.
Initiated2019-10-11
Classified2020-01-24
Terminated2021-05-03
Recall numberZ-0862-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0862-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84281/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.